• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: JOHNSON & JOHNSON SURGICAL VISION, INC. HEALON 5 PRO; OVDS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

JOHNSON & JOHNSON SURGICAL VISION, INC. HEALON 5 PRO; OVDS Back to Search Results
Model Number TV60ML
Medical Device Problem Code Appropriate Device Problem Term/Code Not Available (3191)
Health Effect - Clinical Code No Code Available (3191)
Date of Event 06/09/2020
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Patient age/ date of birth: unknown as information was not provided.Patient gender: unknown as information was not provided.If implanted, give date: not applicable, as the healon 5 pro is not an implantable device.If explanted, give date: not applicable, as the healon 5 pro is not an implantable device.The intraocular lens (iol) is not returning for evaluation as it was discarded by customer; therefore, a failure analysis of the complaint device cannot be completed.A review of the device history record, complaint trending, and risk documentation for this device will be performed.Upon completion of the review, if there is any further relevant information a supplemental medwatch will be filed.(b)(4).All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
 
Event or Problem Description
It was reported a patient experienced capsular entrapment syndrome, experiencing some level of 2-3 diopter shifts following cataract surgery, requiring an immediate post-operative yttrium-aluminum-garnet (yag) laser procedure for correction.Issue was a result of residual healon 5 pro ophthalmic viscosurgical device (ovd) remaining in the eye, potentially posterior to the intraocular lens (iol), following surgery, even after what the surgeon describes as meticulous irrigation aspiration (ia) removal.The patient is doing fine, and progressing nicely according to the surgeon, and he has made the clinical decision to return to usage of the healon duet pro ovd product as a solution.Product was discarded and is not being returned.No additional information was provided.
 
Additional Manufacturer Narrative
This correction mdr is being submitted to address inaccurate information in section f on the xml version of the previous submission caused by a software glitch in the transmission process.The internal non-conformance numbers for this software issue are nr-0148919 and capa-010215.
 
Additional Manufacturer Narrative
Corrected data: in review, it was noted that sections d1 & d4 of the initial mdr report were populated incorrectly.Therefore, the information has been corrected in this supplemental report.The following fields were updated accordingly: section d1: brand name: healon 5 pro.Section d4: model #: tv60ml.All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HEALON 5 PRO
Common Device Name
OVDS
Manufacturer (Section D)
JOHNSON & JOHNSON SURGICAL VISION, INC.
1700 e st andrew place
santa ana CA 92705
MDR Report Key10365671
Report Number3004750704-2020-00042
Device Sequence Number19594090
Product Code LZP
UDI-Device Identifier05050474636996
UDI-Public(01)05050474636996(17)230131(10)UH31112
Combination Product (Y/N)Y
Initial Reporter StateTX
Initial Reporter CountryUS
PMA/510(K) Number
P810031
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source health professional,user faci
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date (Section B) 11/16/2020
Report Date (Section F) 01/01/2005
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Operator of Device Health Professional
Device Expiration Date01/31/2023
Device Model NumberTV60ML
Device Catalogue Number10240015
Device Lot NumberUH31112
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Date Report to Manufacturer01/10/2005
Initial Date Received by Manufacturer 07/08/2020
Supplement Date Received by Manufacturer10/25/2020
10/19/2020
Initial Report FDA Received Date08/04/2020
Supplement Report FDA Received Date10/29/2020
11/16/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Date Report Sent to FDA01/01/2005
-
-