Model Number 15
Medical Device Problem Code Inappropriate/Inadequate Shock/Stimulation (1574)
Health Effect - Clinical Codes No Patient Involvement (2645); No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 07/15/2020
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Physio-control evaluated the customer¿s device and verified the reported issue.Physio replaced the therapy pcb assembly and the transfer relay assembly, and then observed proper device operation through functional and performance testing.The device was returned to the customer for use.
 
Event or Problem Description
The customer contacted physio-control to report that their device displayed ¿abnormal energy delivery¿ when delivering defibrillation therapy during testing.In this state, the device may not have the ability to deliver appropriate defibrillation therapy.There was no patient use associated with the reported event.
 
Additional Manufacturer Narrative
Physio-control determined that the cause of the reported issue was due to the system pcb assembly, the part had failed.No further root cause could be determined.
 
Event or Problem Description
The customer contacted physio-control to report that their device displayed ¿abnormal energy delivery¿ when delivering defibrillation therapy during testing.In this state, the device may not have the ability to deliver appropriate defibrillation therapy.There was no patient use associated with the reported event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LIFEPAK(R) 15 MONITOR/DEFIBRILLATOR
Common Device Name
AUTOMATED EXTERNAL DEFIBRILLATORS (NON-WEARABLE)
Manufacturer (Section D)
PHYSIO-CONTROL, INC. - 3015876
11811 willows road ne
redmond WA 98052
MDR Report Key10365696
Report Number0003015876-2020-00997
Device Sequence Number19408514
Product Code MKJ
UDI-Device Identifier00883873911600
UDI-Public00883873911600
Combination Product (Y/N)N
Initial Reporter StateMI
Initial Reporter CountryUS
PMA/510(K) Number
K142430
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source user facility
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 10/01/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number15
Device Catalogue Number99577-001258
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/27/2020
Initial Date Received by Manufacturer 07/15/2020
Supplement Date Received by Manufacturer09/25/2020
Initial Report FDA Received Date08/04/2020
Supplement Report FDA Received Date10/01/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-