• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC PUERTO RICO OPERATIONS CO. 530G INSULIN PUMP MMT-551LNAH; ARTIFICIAL PANCREAS DEVICE SYSTEM, THRESHOLD SUSPEND

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC PUERTO RICO OPERATIONS CO. 530G INSULIN PUMP MMT-551LNAH; ARTIFICIAL PANCREAS DEVICE SYSTEM, THRESHOLD SUSPEND Back to Search Results
Model Number MMT-551LNAH
Medical Device Problem Codes Display or Visual Feedback Problem (1184); Mechanical Problem (1384); Patient Device Interaction Problem (4001)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 07/31/2020
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).Currently it is unknown whether or not the device may have caused or contributed to the event as no product has been returned.No conclusion can be drawn at this time.We therefore consider this report complete to the best of our knowledge.
 
Event or Problem Description
Information received by medtronic indicated that the insulin pump screen was cracked and hard to read display.Customer stated insulin pump had more noise and slower when rewinding.No harm requiring medical intervention was reported.The insulin pump will not be returned for analysis.
 
Additional Manufacturer Narrative
Device passed the functional test, including the displacement test, rewind, basic occlusion test, occlusion test, prime/a33 test, excessive no delivery test and delivery accyracy test at 0.08730 inches.The stop current and run current measurement tests are within specification.Device also passed self test, off no power alarm test and a21 error test.Device was monitored for 2 days.No unexpected low battery alarm or unexpected off no power alarm noted during testing.No missing segments or partial display, drive or motor anomaly, unexpected motor noise anomaly, rewind anomaly or audio/beep anomaly noted.The motor was tested outside of the device and passed.No cracked display window or damaged lcd glass noted.Device had minor/deep scratched display window.The test p-cap and reservoir does lock in place in the reservoir compartment.This mdr related to the puerto rico manufacturing site has been assigned a medwatch number from the medtronic minimed northridge site, per variance 5.Medtronic, inc.(medtronic) is submitting this report to comply with 21 c.F.R.Part 803, the medical device reporting regulation.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information in the time allotted and has provided as much information as is available to the company as of the submission date this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employees caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any "defects" or has "malfunctioned".These words are included in the fda 3500a form and are fixed items for selection created by the fda, to categorize the type of event solely for the purpose of reporting pursuant to part 803.Medtronic objects to the use of these words and others like it because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
530G INSULIN PUMP MMT-551LNAH
Common Device Name
ARTIFICIAL PANCREAS DEVICE SYSTEM, THRESHOLD SUSPEND
Manufacturer (Section D)
MEDTRONIC PUERTO RICO OPERATIONS CO.
ceiba norte ind. park #50 road
juncos 00777 -386
MDR Report Key10366432
Report Number2032227-2020-157894
Device Sequence Number2310334
Product Code OZO
UDI-Device Identifier00643169507708
UDI-Public(01)00643169507708
Combination Product (Y/N)N
Initial Reporter StateKY
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source consumer
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 09/24/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Model NumberMMT-551LNAH
Device Catalogue NumberMMT-551LNAH
Device Lot NumberA4551LNAHJ
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/26/2020
Initial Date Received by Manufacturer 07/31/2020
Supplement Date Received by Manufacturer09/23/2020
Initial Report FDA Received Date08/05/2020
Supplement Report FDA Received Date09/24/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age41 YR
Patient Weight54
-
-