| Model Number |
3.0X130MM FIBULA NAIL, RIGHT |
| Health Effect - Clinical Code |
Device Embedded In Tissue or Plaque (3165)
|
| Date of Event |
07/15/2020
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
The contribution of the device to the reported event could not be determined as the device was not returned for evaluation.The root cause of the event could not be determined from the information available and without device evaluation.
|
| |
|
Event or Problem Description
|
|
It was reported during an open reduction internal fixation ankle fracture fibula repair the nail was inserted per technique, the surgeon did not like the position so he tried to remove it.During the removal it broke and the end cap remained in the patient.It was observed the talons of the implant were not open.The doctor replaced it with another nail and it was implanted with no issues but a portion of the first nail still remains in the patient.The procedure was completed.The doctor and the patient are planning to do a revision to remove the first partial nail.
|
| |
|