ST. JUDE MEDICAL, INC INQUIRY ELECTROPHYSIOLOGY CATHETER, INQUIRY AFOCUS II EB CATHETER, 110 CM LEN; CATHETER, ELECTRODE RECORDING, OR PROBE, ELECTRODE RECORDING
|
Back to Search Results |
|
| Model Number |
D087023 |
| Medical Device Problem Code |
Material Split, Cut or Torn (4008)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
05/18/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
One decapolar, spiral loop, inquiry afocusii diagnostic catheter was received for evaluation.One image was also submitted for evaluation.The photograph appeared to show the catheter shaft at the loop torn with the conductor wires exposed.Visual inspection revealed the catheter shaft on the loop was torn, exposing the conductor wires.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed.The cause the torn loop and exposed conductor wires is consistent with damage during use.
|
| |
|
Event or Problem Description
|
|
This report is to advise of an event observed during analysis confirming a torn loop, exposing the conductor wires.
|
| |
|