• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STRYKER ENDOSCOPY-SAN JOSE PKG, PRECISION HD 30 DEG., ARTHROSCOPE, 2.3MM X 72MM, C-MOUNT, SPEEDLOCK, A/C

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER ENDOSCOPY-SAN JOSE PKG, PRECISION HD 30 DEG., ARTHROSCOPE, 2.3MM X 72MM, C-MOUNT, SPEEDLOCK, A/C Back to Search Results
Model Number 0502822030
Medical Device Problem Code Optical Problem (3001)
Health Effect - Clinical Codes No Known Impact Or Consequence To Patient (2692); No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 07/09/2020
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Additional information will be provided once the investigation has been completed.The device manufacturer date is not known at this time.However, should it become available it will be provided in future reports.
 
Event or Problem Description
It was reported that there was foggy image.
 
Additional Manufacturer Narrative
The product was returned for investigation and the reported failure mode was confirmed.The failure mode will be monitored for future reoccurrence.This scope was not received for evaluation at stryker endoscopy.This scope was received at henke for evaluation.Based on the henke service record attached, the reported failure [foggy] was confirmed.According to henke: scope evaluated as a level 3.Tip/fiber damage, heavy residue on distal negative lens, dent.Complaint for ¿foggy/circles¿ has been confirmed probably root cause for this failure is: due to customer use and handling.Manufacture date is not known.
 
Event or Problem Description
It was reported that there was foggy image.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PKG, PRECISION HD 30 DEG., ARTHROSCOPE, 2.3MM X 72MM, C-MOUNT, SPEEDLOCK, A/C
Common Device Name
ARTHROSCOPE
Manufacturer (Section D)
STRYKER ENDOSCOPY-SAN JOSE
5900 optical court
san jose CA 95138
MDR Report Key10368120
Report Number0002936485-2020-00330
Device Sequence Number2854236
Product Code HRX
UDI-Device Identifier07613327060089
UDI-Public07613327060089
Combination Product (Y/N)N
Initial Reporter StateIN
Initial Reporter CountryUS
PMA/510(K) Number
K093677
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source consumer
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 09/21/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number0502822030
Device Catalogue Number0502822030
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/15/2020
Initial Date Received by Manufacturer 07/09/2020
Supplement Date Received by Manufacturer07/09/2020
Initial Report FDA Received Date08/05/2020
Supplement Report FDA Received Date09/21/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-