MPRI CAPSURE EPICARDIAL LEAD; DRUG ELUTING PERMANENT RIGHT VENTRICULAR (RV) Back to Search Results
Model Number 4965-50
Medical Device Problem Codes Failure to Capture (1081); Fracture (1260); High impedance (1291); Capturing Problem (2891)
Health Effect - Clinical Codes No Consequences Or Impact To Patient (2199); No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 07/17/2020
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
It was reported that there was a lead impedance warning on the left ventricular (lv) lead for high undefined impedance.A lead fracture is suspected.The alerts will be programmed off and a plan will be discussed with the physician.The lv lead remains in use.No patient complications have been reported as a result of this event.
 
Event or Problem Description
It was further reported that the lv lead thresholds were variable.
 
Additional Manufacturer Narrative
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Additional Manufacturer Narrative
Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Event or Problem Description
It was further reported that the lv lead was confirmed to be fractured and exhibited loss of capture.The lv lead was inactivated and subsequently partially explanted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CAPSURE EPICARDIAL LEAD
Common Device Name
DRUG ELUTING PERMANENT RIGHT VENTRICULAR (RV)
Manufacturer (Section D)
MPRI
road 149 km 56.3
villalba PR 00766
Manufacturer (Section G)
MPRI
road 149 km 56.3
villalba PR 00766
Manufacturer Contact
gerard torenvliet
8200 coral sea st ne
mounds view, MN 55112
7635269277
MDR Report Key10369064
Report Number2649622-2020-14917
Device Sequence Number11005811
Product Code NVN
UDI-Device Identifier00613994501813
UDI-Public00613994501813
Combination Product (Y/N)N
Initial Reporter StateMI
Initial Reporter CountryUS
PMA/510(K) Number
P950024
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2014
Device Explanted Year2024
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup
Report Date (Section B) 06/12/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Expiration Date05/15/2016
Device Model Number4965-50
Device Catalogue Number4965-50
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Received by Manufacturer 07/17/2020
Supplement Date Received by Manufacturer11/23/2020
05/21/2024
Initial Report FDA Received Date08/05/2020
Supplement Report FDA Received Date11/23/2020
06/12/2024
Date Device Manufactured05/15/2014
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
DTBA1D4 CRTD, 6947M62 LEAD, 5568-53 LEAD; DTBA1D4 CRTD, 6947M62 LEAD, 5568-53 LEAD
Outcome Attributed to Adverse Event Required Intervention; Hospitalization;
Patient Age60 YR
Patient SexMale
Patient Weight95 KG
-
-