Model Number 8PERC |
Device Problem
Defective Component (2292)
|
Patient Problem
Extubate (2402)
|
Event Date 05/14/2020 |
Event Type
Injury
|
Manufacturer Narrative
|
If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
According to the reporter, the trach tube broke free from the cuff and was floating freely in the trachea.The patient had a significant cardiac history including rheumatic heart disease, mechanical mitral valve repair, bioprosthetic target vessel revascularization, on anticoagulation, covid negative, and was admitted for stroke/subdural hematoma.The product was replaced.
|
|
Manufacturer Narrative
|
Evaluation summary: medtronic conducted an investigation based upon all information received.A different device was returned than was originally reported, but the device received had the lot number originally reported.Visual inspection noted the flange was detached from the cannula.It was noticed the flange, the right and left hole of the cannula were damaged.A dimensional test was performed to the outer cannula holes and flange lug pins.All the readings were according to specification.It was reported that the cuff seal was defective.The reported issue was confirmed.The most likely cause could not be established from the information available.The manufacturing records for each device are thoroughly reviewed prior to release to ensure that it meets all medtronic quality specifications.A secondary review of the device history records found no potentially contributing factors.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|
Search Alerts/Recalls
|
|
|