(b)(4).Date sent to the fda: 9/8/2020.Additional information was requested and the following was obtained: product code z508g; lot number: pjz889; qty of product involved: (b)(4).A manufacturing record evaluation was performed for the finished device pjz889 batch number, and no non-conformances were identified.Additional h3 investigation summary: it was reported performance breakage suture.An opened sample of product code z508, lot # pjz889 was returned for analysis.During the visual inspection of the open sample, the swage and attachment area were noted to be as expected.The suture was received dispensed from the winding former and no defects, damaged or suture breakage were noted.A functional test was performed and the tensile strength force was above the minimum requirement.The manufacturing records were reviewed, and the manufacturing/packaging criteria were met prior to the release of this batch.Per the condition of the sample, no performance - breakage suture was found and the tested sample met the finished goods requirements.An opened box with three unopened samples of product code z508, lot # pjz899 were returned for analysis.The complaint sample was not received for evaluation.During the visual inspection of three unopened samples, no defects were found on the packages.The samples were opened, and the swage and attachment area were noted to be as expected.The sutures were dispensed without problems and examined along of the strand and no defects, damaged or suture breakage were noted.A functional test was performed and the tensile strength force was above the minimum requirement.The manufacturing records were reviewed, and the manufacturing/packaging criteria were met prior to the release of this batch.Per the condition of samples received, no performance - breakage suture was found and the tested samples met the finished goods requirements.This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon, or its employees that the report constitutes an admission that the product, ethicon, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
|