A device history record (dhr) did not find any defects or nonconformities.All records indicate that the device was manufactured in accordance with all applicable procedures.The root cause of the incident is unknown.Probable cause of the reported incident: since the distal end and bending section rubber were damaged, it is assumed that an external force was applied to the tip, resulting in damage.Due to the above damage, it is presumed that the glue peeling off the bending section rubber caused the leak to occur.In addition, since the date of manufacturing of the actual product is july 31, 2015, the damage of the tip part may be due to aging degradation.The suggested event is preventable in accordance with instruction for use: the workflow of preparation and inspection is shown below.Before each case, prepare and inspect this endoscope as instructed below.Inspect other equipment to be used with this endoscope as instructed in their respective instruction manuals.Using an endoscope that is not functioning properly may compromise patient or operator safety and may result in more severe equipment damage.
|