Model Number 98814-65
Medical Device Problem Code Incorrect, Inadequate or Imprecise Result or Readings (1535)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 07/13/2020
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
No product has been returned.Extended investigation has been performed for the reported complaint and there was no indication that the product did not meet specification.Dhrs for the precision xceed meter were reviewed and the dhrs showed the precision xceed passed all tests prior to release.Dhrs for the precision test strips were reviewed and the dhrs showed the precision test strips passed all tests prior to release.Retain testing was performed for the precision strips and all units performed within specification.If the product is returned, the case will be re-opened and a physical investigation will be performed.All pertinent information available to abbott diabetes care has been submitted.
 
Event or Problem Description
Customer reported receiving erratic readings on their adc blood glucose meter.Customer reported receiving readings of 300 mg/dl and 71 mg/dl within 10 minutes.The results when plotted on a parkes error grid fell into the "c" zone showing the difference in values to be clinically significant.There was no report of death, serious injury, or mistreatment associated with this event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PRECISION XTRA
Common Device Name
BLOOD GLUCOSE MONITORING SYSTEM
Manufacturer (Section D)
ABBOTT DIABETES CARE INC
1360 south loop road
alameda CA 94502 7001
MDR Report Key10370427
Report Number2954323-2020-05281
Device Sequence Number16917627
Product Code NBW
UDI-Device Identifier00093815996958
UDI-Public00093815996958
Combination Product (Y/N)N
Initial Reporter StateTX
Initial Reporter CountryUS
PMA/510(K) Number
K040814
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source consumer
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date (Section B) 08/05/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Expiration Date06/30/2021
Device Model Number98814-65
Device Catalogue Number99695
Device Lot Number4500183017
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/17/2020
Initial Report FDA Received Date08/05/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age55 YR
Patient Weight98
-
-