| Model Number |
3425-0360S |
| Medical Device Problem Codes |
Tear, Rip or Hole in Device Packaging (2385); Device Contaminated During Manufacture or Shipping (2969)
|
| Health Effect - Clinical Codes |
No Known Impact Or Consequence To Patient (2692); No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
07/10/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Device will not be returned.If additional information becomes available, it will be provided in a supplemental report.Device was implanted.Only packaging is available.
|
| |
|
Event or Problem Description
|
|
As reported: "the physician found plastic powder on the device, when the packaging was opened." additionally reported: "the physician cleaned powder on the device and used it." "the sponge [foam of the packaging] was worn.".
|
| |
|
Additional Manufacturer Narrative
|
|
The reported event could not be confirmed, since the implant in question was not returned for evaluation and no other evidence was provided.More detailed information about the complaint event as well as the affected device must be available in order to determine the root cause of the complaint event.The pollution could have been best presented in original condition, which would have been possible in this case.It could not be excluded that the alleged pollution occurred during unpacking the device at user site, which is supported by the evident signs of damage verified at the foam protector.A review of the device history for the reported lot did not indicate any abnormalities.No corrective actions are required at this time.A review of the labeling did not indicate any abnormalities.
|
| |
|
Event or Problem Description
|
|
As reported: "the physician found plastic powder on the device, when the packaging was opened." additionally reported: - "the physician cleaned powder on the device and used it." - "the sponge [foam of the packaging] was worn".
|
| |
|