SMITH & NEPHEW, INC. GII PS HIGH FLEX IMPCTR HD 1-2; PIN, FIXATION, THREADED Back to Search Results
Model Number 71441553
Device Problem Break (1069)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 07/13/2020
Event Type  malfunction  
Event Description
It was reported that before procedure the surgeon stated that the hi-flex impactor head size 1-2 broke and all pieces were recovered.No injuries or surgical delays were reported.An s+n backup device was available.
 
Manufacturer Narrative
The device, intended for use in treatment, was returned for evaluation.A visual inspection confirms the device fractured into three pieces.All pieces were returned for evaluation.The device shows significant signs of wear/usage.The device was manufactured in 2009.A review of complaint history on the listed part revealed no prior complaints for the listed batch with the same failure mode.A review of the manufacturing records did not reveal any manufacturing or material abnormalities that could have caused or contributed to the reported incident.The device is a reusable instrument that can be exposed to numerous surgeries and cleaning cycles.As plastics are vulnerable and crack may have initiated during use and possible causes could be due to the heating and cooling associated with autoclaving or prolonged use.We recommend that all reusable instruments be routinely inspected for wear and damage and replaced as necessary.At this time, we have no reason to suspect that the product failed to meet any product specifications at the time of manufacture.Based on this investigation, the need for corrective action is not indicated.Should additional information be received, the complaint will be reopened.No further investigation warranted for this complaint; however we will continue to monitor for future complaints and investigate as necessary.We consider this investigation closed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GII PS HIGH FLEX IMPCTR HD 1-2
Type of Device
PIN, FIXATION, THREADED
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
MDR Report Key10371216
MDR Text Key201825030
Report Number1020279-2020-03744
Device Sequence Number1
Product Code JDW
UDI-Device Identifier03596010497369
UDI-Public03596010497369
Combination Product (y/n)N
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Type of Report Initial,Followup
Report Date 08/27/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/05/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number71441553
Device Catalogue Number71441553
Device Lot Number09DM05049
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/04/2020
Date Manufacturer Received08/24/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-