Information received by medtronic indicated that the insulin pump had blank display screen.The customer also stated that, the display did not return after pump restart.The customer also statwed that, the insulin pump had unexpected restart alarm.No harm requiring medical intervention was reported.The insulin pump will be returned for analysis.
|
|
Device monitored for several days and no blank display noted.Device received with all operating currents within specification and passed a21 error test.No unexpected a21 alarm anomalies noted.No unexpected audio or vibrate alarm anomalies noted.This mdr related to the puerto rico manufacturing site has been assigned a medwatch number from the medtronic minimed northridge site, per variance 5.Medtronic, inc.(medtronic) is submitting this report to comply with 21 c.F.R.Part 803, the medical device reporting regulation.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information in the time allotted and has provided as much information as is available to the company as of the submission date this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employees caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any "defects" or has "malfunctioned".These words are included in the fda 3500a form and are fixed items for selection created by the fda, to categorize the type of event solely for the purpose of reporting pursuant to part 803.Medtronic objects to the use of these words and others like it because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.
|