| Model Number |
MMT-1780KP |
| Medical Device Problem Code |
No Display/Image (1183)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
08/01/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).Currently it is unknown whether or not the device may have caused or contributed to the event as no product has been returned.The device will be returned for analysis and further information will follow once the analysis has been completed.No conclusion can be drawn at this time.
|
| |
|
Event or Problem Description
|
|
Information received by medtronic indicated that the insulin pump had blank display.Customer stated that the display was not blank less than 30 seconds and did not return.Customer reported that display was not blank currently.Insert battery alarm did not display on the insulin pump screen.Customer stated that the display did not return after insulin pump restart.No harm requiring medical intervention was reported.The device will be returned for analysis.
|
| |
|
Additional Manufacturer Narrative
|
|
Device received with blank display due to disconnected j7 connector.Unable to perform the displacement test, active current test, sleep current test and self test due to blank display.Connector plugged back in and device booted up successfully.
|
| |
|