| Model Number |
MMT-751LNAS |
| Medical Device Problem Codes |
Failure to Cycle (1142); Detachment of Device or Device Component (2907)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
08/01/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).Currently it is unknown whether or not the device may have caused or contributed to the event as no product has been returned.The device will be returned for analysis and further information will follow once the analysis has been completed.No conclusion can be drawn at this time.
|
| |
|
Event or Problem Description
|
|
Information received by medtronic indicated that the insulin pump had compromised force sensor system alarm.Customer stated that the drive support cap was loose.No harm requiring medical intervention was reported.The insulin pump will be returned for analysis.
|
| |
|
Additional Manufacturer Narrative
|
|
Device received with missing segments due to cracked zebra connector.Unable to perform functional testing including the displacement, rewind, basic occlusion, occlusion, prime/a33 and excessive no delivery test or verify a33 alarm due to missing segment.Device received with loose drive support disk, broken belt clip slot, cracked battery tube threads and cracked reservoir tube lip.
|
| |
|