| Model Number |
MMT-723NAS |
| Medical Device Problem Code |
Obstruction of Flow (2423)
|
| Health Effect - Clinical Code |
Hyperglycemia (1905)
|
| Date of Event |
07/29/2020
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
(b)(4).Currently it is unknown whether or not the device may have caused or contributed to the event as no product has been returned.The device will be returned for analysis and further information will follow once the analysis has been completed.No conclusion can be drawn at this time.
|
| |
|
Event or Problem Description
|
|
The customer reported via phone call that the customer had high blood glucose level of 452 mg/dl.Customer was treated with manual injection.Customer stated that the insulin pump had insulin flow block alarm.Customer stated that the insulin pump had motor error alarm.Customer was able to clear alarm.The insulin pump will not be returned for analysis.
|
| |
|
Additional Manufacturer Narrative
|
|
Device passed the displacement test, rewind, basic occlusion, occlusion, prime/a33 and excessive no delivery test.No excessive no delivery/occlusion alarm noted.(b)(4).
|
| |
|