Catalog Number CX*RX15RW30
Medical Device Problem Code Tear, Rip or Hole in Device Packaging (2385)
Health Effect - Clinical Code No Information (3190)
Date of Event 07/16/2020
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The actual sample has been received for evaluation.Visual inspection revealed that the sterile was unopened and there was an area where plastic material had been damaged; stretched.Magnifying inspection found that the stretched area of the sterile bag was where the water port of the oxygenator had been located.In addition, the water port had been deformed.Magnifying inspection of the stretched area revealed that it had been stretched from the inside of the sterile bag.The thickness of the sterile bag material was measured and confirmed to be equivalent to that of a manufacturer retention sample.A review of the device history record and the product release judgement record of the involved product code/lot number combination revealed no findings.Ifu states: do not use if the package or device is damaged (e.G.Cracked) or any of the port caps are off.Based on the provided information and investigation results, there is no definitive evidence that this event was related to a device defect or malfunction.It is likely that since the unit box was deformed, it is likely that an excessive load was applied while the actual sample was in the unit box, causing a hole at the point where the reservoir was in contact with the sterile bag.It is likely to be due to handling during storage period why the actual sample was subjected to such an excessive load.However, the exact cause of the reported event cannot be definitively determined based on the available information.
 
Event or Problem Description
The user facility reported that the involved capiox device was used pre-treatment.After they opened the box, the perfusionist found aseptic package broken.The procedure outcome and patient outcome was not reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CAPIOX RX15
Common Device Name
OXYGENATOR, CARDIOPULMONARY BYPASS
Manufacturer (Section D)
TERUMO CORPORATION, ASHITAKA
150 maimaigi-cho
fujinomiya city, 418
JA  418
MDR Report Key10372187
Report Number9681834-2020-00154
Device Sequence Number14314283
Product Code DTZ
UDI-Device Identifier04987350771056
UDI-Public04987350771056
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
K051997
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date (Section B) 08/06/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date09/30/2022
Device Catalogue NumberCX*RX15RW30
Device Lot Number191010C
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/27/2020
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/17/2020
Initial Report FDA Received Date08/06/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age7 YR
Patient Weight22
-
-