• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC MED REL MEDTRONIC PUERTO RICO PROTECTA XT CRT-D; DEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER, WITH CARDIAC RESYNCHRONIZATIO

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC MED REL MEDTRONIC PUERTO RICO PROTECTA XT CRT-D; DEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER, WITH CARDIAC RESYNCHRONIZATIO Back to Search Results
Model Number D314TRG
Medical Device Problem Code Battery Problem (2885)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 06/20/2018
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
It was reported that the cardiac resynchronization therapy defibrillator (crt-d) had not been working properly and that the battery had already become weak.The device was explanted and replaced.No patient complications have been reported as a result of this event.
 
Additional Manufacturer Narrative
Product event summary: the device was returned and analyzed.Analysis of the device revealed normal battery depletion.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PROTECTA XT CRT-D
Common Device Name
DEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER, WITH CARDIAC RESYNCHRONIZATIO
Manufacturer (Section D)
MEDTRONIC MED REL MEDTRONIC PUERTO RICO
road 31, km. 24, hm 4
ceiba norte industrial park
juncos PR 00777
MDR Report Key10372295
Report Number3004209178-2020-13532
Device Sequence Number2567262
Product Code NIK
UDI-Device Identifier00613994677884
UDI-Public00613994677884
Combination Product (Y/N)N
Initial Reporter StateNC
Initial Reporter CountryUS
PMA/510(K) Number
P010031
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2012
Device Explanted Year2018
Reporter Type Manufacturer
Report Source consumer
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 09/01/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Expiration Date12/28/2013
Device Model NumberD314TRG
Device Catalogue NumberD314TRG
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/28/2018
Initial Date Received by Manufacturer 07/14/2020
Supplement Date Received by Manufacturer08/19/2020
Initial Report FDA Received Date08/06/2020
Supplement Report FDA Received Date09/01/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
5076-52 LEAD, 694965 LEAD, 419488 LEAD; 5076-52 LEAD, 694965 LEAD, 419488 LEAD
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
Patient Age81 YR
-
-