| Model Number |
MMT-723NAL |
| Medical Device Problem Code |
No Display/Image (1183)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
07/30/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).Currently it is unknown whether or not the device may have caused or contributed to the event as no product has been returned.No conclusion can be drawn at this time.We therefore consider this report complete to the best of our knowledge.
|
| |
|
Event or Problem Description
|
|
Information received by medtronic indicated that they insulin pump had blank display.No harm requiring medical intervention was reported.Customer stated that insulin pump was not exposed to moisture.The customer was assisted with troubleshooting and customer reports display was blank currently.Customer stated the display was not blank less than 30 seconds and did not returned.The customer removes the battery and stated the insert battery alarm did not display on the insulin pump screen.Customer stated the contact on the battery cap was neither missing nor damage.Customer stated battery compartment was neither damaged nor corroded.Customer stated the spring was neither damaged nor corroded.Customer stated that display did not returned.The insulin pump will not be returned for analysis.
|
| |
|