MAUDE Adverse Event Report: DAVOL INC., SUB. C.R. BARD, INC. SIMPULSE PLUS SUCTION IRRIGATOR TIP; LAVAGE, JET

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DAVOL INC., SUB. C.R. BARD, INC. SIMPULSE PLUS SUCTION IRRIGATOR TIP; LAVAGE, JET Back to Search Results
Catalog Number 0057621
Medical Device Problem Codes Device Contamination with Chemical or Other Material (2944); Manufacturing, Packaging or Shipping Problem (2975)
Health Effect - Clinical Code No Patient Involvement (2645)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The subject product was not returned; however, a photo was provided.Photo evaluation confirms the presence of an unidentified particulate within the package.Review of the photo was not helpful in determining the source of the foreign material.A determination of what the foreign material is cannot be made without the sample to evaluate, as such, no conclusion can be made.A review of the manufacturing records was performed for the subject lot and found that the lot was manufactured to the specification.If/when additional information is provided, a supplemental emdr will be submitted.Not returned - sample unavailable.
 
Event or Problem Description
It was reported that during inspection at the hospital warehouse, an unknown particulate was observed inside the simpulse plus showerhead tip suction irrigator package.There was no patient involvement as the reported product issue was discovered during inspection.
 
Additional Manufacturer Narrative
The subject product was not returned; however, a photo was provided.Photo evaluation confirms the presence of an unidentified particulate within the package.Review of the photo was not helpful in determining the source of the foreign material.A determination of what the foreign material is cannot be made without the sample to evaluate, as such, no conclusion can be made.A review of the manufacturing records was performed for the subject lot and found that the lot was manufactured to the specification.Addendum: h.11: this is an addendum to the initial emdr.This supplemental emdr was submitted to report the corrected investigation findings.Based on the investigation activities performed, the reported event was confirmed.Although the provided photos confirm for foreign matter inside the sealed packages, the origin of the foreign matter could not be determined.However, as the foreign matter was present in the package prior to sealing, this presented at the time of manufacturing and is confirmed for a manufacturing related condition.Awareness training was provided to appropriate associates.Should additional information be provided, a supplemental emdr will be submitted.Note: section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.H3 other text : not returned - sample unavailable.
 
Event or Problem Description
It was reported that during inspection at the hospital warehouse, an unknown particulate was observed inside the simpulse plus showerhead tip suction irrigator package.There was no patient involvement as the reported product issue was discovered during inspection.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SIMPULSE PLUS SUCTION IRRIGATOR TIP
Common Device Name
LAVAGE, JET
Manufacturer (Section D)
DAVOL INC., SUB. C.R. BARD, INC.
100 crossings blvd.
warwick RI 02886
MDR Report Key10372555
Report Number1213643-2020-07154
Device Sequence Number2925621
Product Code FQH
UDI-Device Identifier00801741010415
UDI-Public(01)00801741010415
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
K830386
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,distri
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 08/20/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Other
Device Expiration Date06/28/2024
Device Catalogue Number0057621
Device Lot NumberJUDUF033
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 07/14/2020
Supplement Date Received by Manufacturer08/12/2020
Initial Report FDA Received Date08/06/2020
Supplement Report FDA Received Date08/20/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-