Model Number N/A
Medical Device Problem Code Fracture (1260)
Health Effect - Clinical Code No Information (3190)
Date of Event 02/10/2020
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
(b)(4).Medical product: vgd ps open por fem rt 62.5 catalog # 184506 lot # 592860; bmet regenx pri tib tray 67mm catalog # 141272 lot # 937420; biomet finned pri stem 40mm catalog # 141314 lot # 342620; e1 vngd ps tib brg 63/67x10 catalog # ep-183620 lot # 587500.Customer has indicated that the product will not be returned because product location is unknown.Reported event was confirmed by review of medical records provided.Device was not returned.Device history record (dhr) was reviewed and no discrepancies relevant to the reported event were found.The root cause of the patella peg fracture was determined to be a labeling and training deficiency as the surgical technique addendum for installation of regenerex patella failed to include guidance on depth line indication on the peg drill bit.The event is being addressed through the capa process.Medical records review indicates that two patella pegs were fractured.There was no synovitis or metallosis seen.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event or Problem Description
It was reported that the patient underwent a knee arthroplasty.Subsequently, the patient was revised due to two fractured pegs on the patella prosthesis.The patella was removed and replaced.
 
Additional Manufacturer Narrative
H6- type of investigation - device not return is n/a which was reported on initial report.Complaint sample was evaluated and the reported event was confirmed.Visual evaluation of the returned device shows wear/discoloration / foreign material and two of the posts have fractured.The additional information received doesn't change the previous investigation.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event or Problem Description
No further event information available at the time of this report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
REGENERX 3 PEG SER A PATELLA 28MM
Common Device Name
PROSTHESIS, KNEE
Manufacturer (Section D)
ZIMMER BIOMET, INC.
56 e. bell drive
warsaw IN 46582
MDR Report Key10372592
Report Number0001825034-2020-03017
Device Sequence Number16875206
Product Code OIY
UDI-Device Identifier00880304554405
UDI-Public(01)00880304554405
Combination Product (Y/N)N
Initial Reporter StateGA
Initial Reporter CountryUS
PMA/510(K) Number
K171054
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2015
Device Explanted Year2020
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source consumer,distributor,health p
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 10/06/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date09/08/2020
Device Model NumberN/A
Device Catalogue Number141355
Device Lot Number044830
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/10/2021
Was the Report Sent to FDA? No
Initial Date Received by Manufacturer 07/15/2020
Supplement Date Received by Manufacturer10/04/2021
Initial Report FDA Received Date08/06/2020
Supplement Report FDA Received Date10/07/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
N/A
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
SEE H10 NARRATIVE
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
-
-