• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DAVOL INC., SUB. C.R. BARD, INC. -1213643 MESH ¿ COMPOSIX L/P; SURGICAL MESH

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DAVOL INC., SUB. C.R. BARD, INC. -1213643 MESH ¿ COMPOSIX L/P; SURGICAL MESH Back to Search Results
Catalog Number 0134460
Medical Device Problem Codes Defective Device (2588); Patient-Device Incompatibility (2682); Material Deformation (2976); Insufficient Device Problem Information (3190); Migration (4003)
Health Effect - Clinical Codes Adhesion(s) (1695); Granuloma (1876); Pain (1994); Scar Tissue (2060); Hernia (2240); Injury (2348); Disability (2371); Obstruction/Occlusion (2422)
Date of Event 07/23/2013
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
No conclusions can be made.The patient's attorney alleges "past, present, and future damages, including but not limited to, mental and physical pain and suffering for severe and permanent personal injuries sustained by the patient"; however, no details have been provided.No lot number has been provided; therefore, a review of the manufacturing records is not possible.Should additional information be provided, a supplemental emdr will be submitted.Not returned.
 
Event or Problem Description
Attorney alleges that the patient underwent surgery for implant of unspecified bard/davol composix l/p and xenmatrix surgical graft on (b)(6) 2011 and/or (b)(6) 2013.As reported, the patient is making a claim for an adverse patient outcome against the composix l/p.Attorney alleges general allegations for "past, present, and future damages, including but not limited to, mental and physical pain and suffering for severe and permanent personal injuries sustained by the patient." it is also alleged that the patient experienced emotional distress and the device was defective.
 
Additional Manufacturer Narrative
No conclusions can be made.The patient's attorney alleges "past, present, and future damages, including but not limited to, mental and physical pain and suffering for severe and permanent personal injuries sustained by the patient"; however, no details have been provided.No lot number has been provided; therefore, a review of the manufacturing records is not possible.Addendum: this supplemental emdr is submitted to document additional information provided.Root cause is inconclusive; no conclusion can be made as to the extent to which the implant may have caused or contributed to the patient's postoperative course.Note, the manufacturing date (15-jan-2011) is considered to be a best estimate.Note: section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
 
Event or Problem Description
Attorney alleges that the patient underwent surgery for implant of unspecified bard/davol composix l/p and xenmatrix surgical graft on (b)(6) 2011 and/or (b)(6) 2013.As reported, the patient is making a claim for an adverse patient outcome against the composix l/p.Attorney alleges general allegations for "past, present, and future damages, including but not limited to, mental and physical pain and suffering for severe and permanent personal injuries sustained by the patient." it is also alleged that the patient experienced emotional distress and the device was defective.Addendum per additional information provided: (b)(6) 2011 - patient was diagnosed with internal hernia, incisional hernia thereby underwent laparoscopic repair with implant of composix l/p mesh (device #1).Per op notes, "a composix l/p mesh (device #1) was placed and sutured." (b)(6) 2012 - patient was diagnosed with ventral hernia, small bowel obstruction thereby underwent open repair.Per op notes, "adhesions were taken down and a primary repair was done using sutures." (b)(6) 2013 - patient was diagnosed with multiple recurrent incisional hernias, adhesions thereby underwent laparoscopic repair.Per op notes, "adhesions were taken for the previous mesh (device #1).Th mesh (device #1) appeared to be contracted and pulled away from the tacks." (b)(6) 2013 - patient was diagnosed with multiple recurrent incisional hernias and mesh granuloma along with keloid scar thereby underwent open repair with explant of composix l/p mesh (device #1) and implant of xenmatrix (device #2).Per op notes, "the mesh granuloma was encountered.The contracted mesh (device #1) was excised along with tacks and removed.A xenmatrix (device #2) was placed and sutured.".
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MESH ¿ COMPOSIX L/P
Common Device Name
SURGICAL MESH
Manufacturer (Section D)
DAVOL INC., SUB. C.R. BARD, INC. -1213643
100 crossings blvd.
warwick RI 02886
Manufacturer (Section G)
BARD SHANNON LIMITED 3005636544
san geronimo industrial park
lot #1, road #3, km 79.7
humacao PR 00791
Manufacturer Contact
andrew topoulos
100 crossings blvd.
warwick, RI 02886
4018258495
MDR Report Key10372777
Report Number1213643-2020-07157
Device Sequence Number16875213
Product Code FTL
UDI-Device Identifier00801741031274
UDI-Public(01)00801741031274(17)160228(10)HUVA1708
Combination Product (Y/N)N
Initial Reporter StateRI
Initial Reporter CountryUS
PMA/510(K) Number
K061754
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2011
Device Explanted Year2013
Reporter Type Manufacturer
Report Source Other,Consumer
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 05/26/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date02/28/2016
Device Catalogue Number0134460
Device Lot NumberHUVA1708
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/03/2020
Supplement Date Received by Manufacturer05/20/2025
Initial Report FDA Received Date08/06/2020
Supplement Report FDA Received Date05/26/2025
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured01/15/2011
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Disability; Required Intervention;
Patient Age43 YR
Patient SexFemale
-
-