| Model Number |
7210569F |
| Medical Device Problem Codes |
Break (1069); Positioning Problem (3009)
|
| Health Effect - Clinical Code |
No Patient Involvement (2645)
|
| Date of Event |
07/09/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that the handle of the spider limb broke off, which caused that the device could not be used since there is no way to secure the position of the patient without this handle.Incident occurred before the shoulder repair procedure.There was no delays and a back-up device was available.No patient injuries were reported.All available information has been disclosed.If additional information should become available, a supplemental report will be submitted accordingly.
|
| |
|
Additional Manufacturer Narrative
|
|
H10 h6: the device, intended for use in treatment, was received for evaluation.There was a relationship found between the returned device and the reported incident.A visual evaluation revealed that the rotational handle was bent.A functional evaluation revealed that the unit could not be properly attached to the rail clamp because of the bent rotational handle.The complaint was confirmed.Factors that could have contributed to the reported event include an impact event or application of excessive torque inconsistent with intended use.A review of the device history record shows there are no indications to suggest that the product did not meet manufacturing specification or would not be able to perform as intended.A complaint history review concluded this was a repeat issue.
|
| |
|