| Model Number |
MMT-551LNAP |
| Medical Device Problem Codes |
Display or Visual Feedback Problem (1184); Mechanical Problem (1384); Device Difficult to Program or Calibrate (1496); Obstruction of Flow (2423)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
07/30/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).Currently it is unknown whether or not the device may have caused or contributed to the event as no product has been returned.No conclusion can be drawn at this time.We therefore consider this report complete to the best of our knowledge.
|
| |
|
Event or Problem Description
|
|
Information received by medtronic indicated that the insulin pump buttons were sticking.Customer stated that the insulin pump had no delivery alarm.Customer stated that the motor had sluggish and louder during rewind.Customer stated that the insulin shoots out during priming.No harm requiring medical intervention was reported.The insulin pump will be returned for analysis.
|
| |
|