Model Number OER-PRO |
Device Problem
Device Contamination with Chemical or Other Material (2944)
|
Patient Problem
No Patient Involvement (2645)
|
Event Date 06/01/2020 |
Event Type
malfunction
|
Manufacturer Narrative
|
The reported problem was resolved through an onsite service visit by an olympus field service engineer (fse).The fse inspected all disinfectant hoses for evidence of degradation; tank hose 1, tank hose 2 and the disinfectant hose were replaced due to degradation.
|
|
Event Description
|
A user facility reported to olympus that a black substance was observed in the mesh filters inside the basin.There was no patient injury or harm, associated with the problem, reported to olympus.
|
|
Manufacturer Narrative
|
Upon further review this is not a reportable malfunction.Per the legal manufacturer there is no potential for this issue to cause or contribute to death or serious injury if the malfunction were to recur.
|
|