| Model Number |
1012453-20 |
| Medical Device Problem Codes |
Deflation Problem (1149); Entrapment of Device (1212)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
07/13/2020
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
The device is expected to be returned for evaluation.It has not yet been received.A follow up report will be submitted with all additional relevant information.The additional nc trek device is filed under a separate medwatch report number.
|
| |
|
Event or Problem Description
|
|
It was reported that the procedure was to treat a lesion in the right coronary artery.The 4.00x20mm nc trek balloon dilatation catheter (bdc) was prepped outside the anatomy and the contrast mix ratio was 50/50.The bdc was inflated once to dilate the artery.After inflation, the balloon failed to deflate and could not be retrieved into the guiding catheter and could not advance further in the radial artery.A surgeon was called and a cut down was performed to retrieve the balloon from the radial artery.A second 4.00x20mm nc trek bdc was also prepped outside the anatomy and the contrast mix ratio was 50/50.The bdc was advanced through a 6f guiding catheter through access on the other side via the left radial artery.The bdc was inflated five times at 20 atmospheres.Although negative pressure was held for 15 seconds, the bdc only partially deflated.The bdc was able to be retracted into the guiding catheter and both were removed from the patient anatomy as a single unit.The procedure was successfully completed with the patient in good condition.There were no adverse patient sequela and a procedural delay of three hours occurred due to the surgical intervention for the first trek balloon.No additional information was provided.
|
| |
|
Event or Problem Description
|
|
Subsequent to the initially filed report: the 4.00x20mm nc trek bdc was cut at the proximal end of the hypotube and hub in order to be able to remove the device from the patient anatomy via a cutdown.No additional information was provided.
|
| |
|
Additional Manufacturer Narrative
|
|
The device was returned for analysis.The reported deflation problem was unable to be replicated in a testing environment due to the condition of the returned device.The reported entrapment of device was unable to be replicated in a testing environment as it was based on operational circumstances.A review of the lot history record identified one manufacturing nonconformities issued to the reported lot that would have contributed to this event.Additionally, a review of the complaint history identified two other similar incidents from this lot.One sterile/unused device part 1012453-20/ lot 00121g1 was returned.The balloon was inflated and deflated three times meeting specification.The balloon folded flat all three times.The complaint investigation determined the reported deflation problem is related to a potential manufacturing issue associated with deflation issues.The reported entrapment of device was a direct symptom of the reported deflation issue.The issue is being addressed per internal operating procedures.Abbott vascular will continue to trend the performance of these devices.
|
| |
|