| Model Number |
N/A |
| Medical Device Problem Codes |
Positioning Failure (1158); Difficult to Insert (1316)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
06/25/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).Medical product: tibia cemented 5 degree stemmed right size j, item#: 42532008802, lot# : 63886950.Report source - (b)(6).Customer has indicated that the product will be returned to zimmer biomet for investigation.The investigation is in process.Once the investigation has been completed, a follow-up mdr will be submitted.
|
| |
|
Event or Problem Description
|
|
It was reported that an articular surface would not seat with the tibial tray.Upon insertion of the articular surface, the surgeon felt insertion difficulty so he tried to insert it into tibia component with an impactor but it was deformed.So he decided to use an 11mm surface to complete the procedure.There is no additional information available at this time.
|
| |
|
Additional Manufacturer Narrative
|
|
This follow-up report is being submitted to relay additional information.Complaint is confirmed.Visual inspection of the returned device exhibits flared dovetail.Articular surface is compatible with the tibia, however entire construct compatibility could not be assessed due to insufficient information.Review of the device history record identified no related deviations or anomalies during manufacturing.Root cause cannot be determined.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|
| |
|
Event or Problem Description
|
|
No further event information is available at the time of this report.
|
| |
|
Search Alerts/Recalls
|