| Model Number |
ME2017 |
| Medical Device Problem Code |
Insufficient Device Problem Information (3190)
|
| Health Effect - Clinical Code |
No Information (3190)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
An unexpected return confirmed a male luer break.Visual inspection as received confirmed the male luer collar to be cracked and broken.Closer inspection of the break under a lab microscope observed no stress marks or tool marks.The set was visually inspected for cracks, misassemblies or damages to the components.Visual inspection found that a male luer collar of the extension set was cracked and broken.Device history record for model me2017 could not be performed due to no lot number being provided by customer.The root cause was identified as related to design of the specific male luer component.
|
| |
|
Event or Problem Description
|
|
An unexpected tubing set with an unspecified issue was returned by the customer.No further information is available.
|
| |
|