• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: CAREFUSION ALARIS EXTENSION SET; SET, EXTENSION, INTRAVASCULAR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

CAREFUSION ALARIS EXTENSION SET; SET, EXTENSION, INTRAVASCULAR Back to Search Results
Model Number 10010570
Medical Device Problem Code Insufficient Device Problem Information (3190)
Health Effect - Clinical Code No Information (3190)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
An unexpected return from the customer confirmed a female luer crack.Visual inspection confirmed that the female luer had a lateral crack.Closer inspection under a lab microscope observed no tool marks or signs of over torque.Functional testing confirmed a leak from the observed crack.No other leaks were observed throughout the set.The set was visually inspected for kinks, incomplete bonding engagements, holes/tears in the tubing or damages to the components.Visual inspection found a crack in the female luer wall on the opposite end of the injection gate of the set.The location of the luer gate on the female luer (at the ¿wings¿) indicates that this female luer was manufactured before the change order that moved the injection gate to a lower location to help mitigate and prevent female luer cracks.No tool marks or signs of over torque were observed.No damages were observed on the male luer.Functional testing resulted to a leakage from a cracked female luer.Device history record for model 10015817 could not be performed due to no lot number being provided by customer.The root cause of the crack is a result of internal stresses created in the luer during the manufacturing process that make it more susceptible to chemical/mechanical attacks.
 
Event or Problem Description
Product was received as an unexpected return with an unspecified failure.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ALARIS EXTENSION SET
Common Device Name
SET, EXTENSION, INTRAVASCULAR
Manufacturer (Section D)
CAREFUSION
10020 pacific mesa blvd
san diego CA 92121 4386
MDR Report Key10376207
Report Number9616066-2020-02450
Device Sequence Number11498582
Product Code FPA
UDI-Device Identifier10885403233500
UDI-Public10885403233500
Combination Product (Y/N)N
Initial Reporter StateMI
Initial Reporter CountryUS
PMA/510(K) Number
K790108
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Nurse
Type of Report Initial
Report Date (Section B) 03/19/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number10010570
Device Catalogue Number10010570
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/19/2020
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer Not provided
Initial Report FDA Received Date08/06/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-