• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: JOHNSON & JOHNSON SURGICAL VISION, INC. TECNIS; TORIC IOLS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

JOHNSON & JOHNSON SURGICAL VISION, INC. TECNIS; TORIC IOLS Back to Search Results
Model Number ZCT150
Medical Device Problem Code Unintended Movement (3026)
Health Effect - Clinical Codes Visual Impairment (2138); No Code Available (3191)
Date of Event 03/10/2020
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
(b)(4).Device evaluation: the complaint lens was received submerged in an unknown solution inside a specimen cup.Additionally, a shipping label was received as well.Visual inspection under magnification revealed a detached haptic and that the lens was received cut in half, which is consistent with a lens that was handled during explant.Dimensional inspection was performed (haptic width and haptic thickness on the remaining haptic), and all measurements were within specifications.Based on the return condition of the lens no further evaluation could be performed.The complaint issue could not be confirmed, and no product deficiency could be identified.Manufacturing record review: the manufacturing process record was evaluated and revealed that the product was manufactured and released according to specifications.A search revealed that no other complaints were received from this production order.Conclusion: as a result of the investigation there is no indication of a product quality deficiency and the reported issue could not be verified.All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
 
Event or Problem Description
It was reported that an intraocular lens (iol) rotated 30 degrees and was noted to manifest as decreased visual acuity.Therefore, the iol was explanted from the patient¿s right eye.It was replaced with a different model iol, 21.5 diopter power.There was no incision enlargement, no vitrectomy, or no sutures required.There was no patient injury post-op.No other information was provided.
 
Additional Manufacturer Narrative
This correction mdr is being submitted to address inaccurate information in section f on the xml version of the previous submission caused by a software glitch in the transmission process.The internal non-conformance numbers for this software issue are nr-0148919 and capa-010215.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TECNIS
Common Device Name
TORIC IOLS
Manufacturer (Section D)
JOHNSON & JOHNSON SURGICAL VISION, INC.
1700 e st andrew place
santa ana CA 92705
MDR Report Key10377035
Report Number9614546-2020-00293
Device Sequence Number16938315
Product Code HQL
UDI-Device Identifier05050474552180
UDI-Public(01)05050474552180(17)200503
Combination Product (Y/N)Y
Initial Reporter StateVA
Initial Reporter CountryUS
PMA/510(K) Number
P980040
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2019
Device Explanted Year2020
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 10/29/2020
Report Date (Section F) 01/01/2005
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date05/03/2020
Device Model NumberZCT150
Device Catalogue NumberZCT150U220
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/29/2020
Was the Report Sent to FDA? No
Date Report to Manufacturer01/10/2005
Initial Date Received by Manufacturer 07/10/2020
Supplement Date Received by Manufacturer10/25/2020
Initial Report FDA Received Date08/06/2020
Supplement Report FDA Received Date10/29/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age73 YR
Date Report Sent to FDA01/01/2005
-
-