| Model Number |
FOS410 |
| Medical Device Problem Code |
Fracture (1260)
|
| Health Effect - Clinical Code |
No Information (3190)
|
| Date of Event |
01/24/2020
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Zimmerbiomet complaint number (b)(4).
|
| |
|
Event or Problem Description
|
|
It was reported implant was removed due to fracture.
|
| |
|
Additional Manufacturer Narrative
|
|
Zimmerbiomet complaint number: (b)(4).The full osseotite® implant 4 x 10mm (fos410) was not returned.Since product has not been returned, visual/functional evaluation could not be performed.The investigation has been performed based on the available information.No pre-existing conditions were noted on the product experience report.The reported product was located on an unknown tooth site and used for approximately 4 months.The reported event could not be recreated due to the nature of the dental device and event (fracture).The customer has not provided additional pictures or x-ray images of the product.Dhr review was completed for the subject lot number 2019060198.It was confirmed that all operations and inspections were executed as per applicable procedure.No deviations or non-conformances, which could have caused or contributed to the reported event was noted as part of the dhr.Lot was inspected and passed all acceptance criteria by qa.Complaint history review was performed for the reported lot number (2019060198) for similar event and no other complaint was identified.August post market trending was reviewed and there were no actionable events or corrective actions for the reported event (implant fracture) or product (fos410).Therefore, based on the available information, device malfunction could not be verified and the reported event was non-verifiable.
|
| |
|
Event or Problem Description
|
|
No further event information is available at the time of this report.
|
| |
|
Additional Manufacturer Narrative
|
|
Zimmerbiomet complaint number (b)(4).The following sections have been updated: b4: date of this report b5: describe event or problem *d3: manufacturer *d4: expiration date udi: (b)(4) *d9: device available for evaluation change ¿no' to 'yes' g2: office contact - manufacturing site g3: date received by manufacturer h1: type of report, follow-up number h2: follow up type h3: device evaluated by manufacturer: change ¿no' to 'yes' h4: device manufacture date h6: evaluation codes h10: additional narrative one full osseotite® implant 4 x 10mm (fos410) was returned for inspection.Inspection of the returned device notes that the implant is fractured across the collar.No pre-existing conditions were noted on the per.The reported product was located on an unknown tooth site and used for approximately 4 months.The reported event could not be recreated due to the nature of the dental device and event (fracture).The customer has not provided additional pictures or x-ray images of the product.Review of appropriate documentation: documents reviewed: p-iis086gi rev.G 10/2019; potential adverse events; page 1 also, according to the ifu, patient factors like presence of occlusal abnormalities or parafunctional habits (e.G.Severe bruxism, clenching, overloading or gnawing) may cause screw loosening, restoration fracture, and/or implant failure.Dhr review was completed for the subject lot number 2019060198.It was confirmed that all operations and inspections were executed as per applicable procedure.No deviations or non-conformances, which could have caused or contributed to the reported event was noted as part of the dhr.Lot was inspected and passed all acceptance criteria by qa.Complaint history review was performed for the reported lot number (2019060198) for similar event and no other complaint was identified.December post market trending was reviewed and there were no actionable events or corrective actions for the reported event (implant fracture) or product (fos410).Therefore, based on the available information, device malfunction has occurred and the reported event was confirmed.
|
| |
|
Event or Problem Description
|
|
No further event information is available at the time of this report.
|
| |
|