MAUDE Adverse Event Report: MEDTRONIC PUERTO RICO OPERATIONS CO. 670G INSULIN PUMP MMT-1780KL; ARTIFICIAL PANCREAS DEVICE SYSTEM, SINGLE HORMONAL CONTROL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC PUERTO RICO OPERATIONS CO. 670G INSULIN PUMP MMT-1780KL; ARTIFICIAL PANCREAS DEVICE SYSTEM, SINGLE HORMONAL CONTROL Back to Search Results
Model Number MMT-1780KL
Medical Device Problem Code Device Difficult to Program or Calibrate (1496)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 07/28/2020
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).Currently it is unknown whether or not the device may have caused or contributed to the event as no product has been returned.The device will be returned for analysis and further information will follow once the analysis has been completed.No conclusion can be drawn at this time.
 
Event or Problem Description
Information received by medtronic indicated that the insulin pump ok button sometimes did not work customer stated that numbers were ramping on their own.Customer stated the scroll bar was moving with no input.Troubleshooting was performed.No harm requiring medical intervention was reported.The insulin pump will be returned for analysis.
 
Additional Manufacturer Narrative
Device passed the displacement test and self test.No keypad anomalies noted during testing.Keypad unresponsive/button error alarm not confirmed.(b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
670G INSULIN PUMP MMT-1780KL
Common Device Name
ARTIFICIAL PANCREAS DEVICE SYSTEM, SINGLE HORMONAL CONTROL
Manufacturer (Section D)
MEDTRONIC PUERTO RICO OPERATIONS CO.
ceiba norte ind. park #50 road
juncos 00777 -386
MDR Report Key10377682
Report Number2032227-2020-159337
Device Sequence Number19714934
Product Code OZP
UDI-Device Identifier00763000365882
UDI-Public(01)00763000365882(17)230620
Combination Product (Y/N)N
Initial Reporter StateSD
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source consumer
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 09/07/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Expiration Date06/20/2023
Device Model NumberMMT-1780KL
Device Catalogue NumberMMT-1780KL
Device Lot NumberHG4EDQU
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/17/2020
Initial Date Received by Manufacturer 08/03/2020
Supplement Date Received by Manufacturer09/04/2020
Initial Report FDA Received Date08/07/2020
Supplement Report FDA Received Date09/08/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age59 YR
Patient Weight122
-
-