| Model Number |
MMT-723RNAS |
| Medical Device Problem Code |
Device Difficult to Program or Calibrate (1496)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
07/26/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).Currently it is unknown whether or not the device may have caused or contributed to the event as no product has been returned.The device will be returned for analysis and further information will follow once the analysis has been completed.No conclusion can be drawn at this time.
|
| |
|
Event or Problem Description
|
|
Information received by medtronic indicated that the up arrow on the insulin pump was not responding and he used fingernail to press it.The time on the insulin pump advanced.No harm requiring medical intervention was reported.The insulin pump will be returned for analysis.
|
| |
|
Additional Manufacturer Narrative
|
|
Device passed displacement test and self test.Device received with intermittent button response due to flattened up and down button dome switch (creased).No button error alarm during testing.No unlocked lcd keypad connector.(b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
Device passed displacement test and self test.Device received with intermittent button response due to flattened up and down button dome switch (creased).No button error alarm during testing.No unlocked j2/lcd keypad connector.
|
| |
|
Search Alerts/Recalls
|