• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC PUERTO RICO OPERATIONS CO. 530G INSULIN PUMP MMT-751LNAH; ARTIFICIAL PANCREAS DEVICE SYSTEM, THRESHOLD SUSPEND

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC PUERTO RICO OPERATIONS CO. 530G INSULIN PUMP MMT-751LNAH; ARTIFICIAL PANCREAS DEVICE SYSTEM, THRESHOLD SUSPEND Back to Search Results
Model Number MMT-751LNAH
Medical Device Problem Code Failure to Cycle (1142)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 08/03/2020
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).Currently it is unknown whether or not the device may have caused or contributed to the event as no product has been returned.The device will be returned for analysis and further information will follow once the analysis has been completed.No conclusion can be drawn at this time.
 
Event or Problem Description
Information received by medtronic indicated that the insulin pump received compromised forced sensor alarm during rewinding and priming.Customer reported the drive support cap was normal.The force sensor was no longer detecting reservoir during seating.No harm requiring medical intervention was noted.The insulin pump will be returned for analysis.
 
Additional Manufacturer Narrative
Device passed the displacement test.Device received with motor error alarm during the basic occlusion test due to loose or protruded drive support disk.Unable to perform prime or a33 test, excessive no delivery test and occlusion test or verify a33 alarm due to motor error alarm.Device received with cracked belt clip slot, missing end cap sticker, stained address or serial number label and cracked reservoir tube lip.The test infusion set and reservoir does lock into place.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
530G INSULIN PUMP MMT-751LNAH
Common Device Name
ARTIFICIAL PANCREAS DEVICE SYSTEM, THRESHOLD SUSPEND
Manufacturer (Section D)
MEDTRONIC PUERTO RICO OPERATIONS CO.
ceiba norte ind. park #50 road
juncos 00777 -386
MDR Report Key10377810
Report Number2032227-2020-159414
Device Sequence Number3167876
Product Code OZO
UDI-Device Identifier00643169507753
UDI-Public(01)00643169507753
Combination Product (Y/N)N
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source other
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 02/10/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Model NumberMMT-751LNAH
Device Catalogue NumberMMT-751LNAH
Device Lot NumberA4751LNAHJ
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/30/2021
Initial Date Received by Manufacturer 08/03/2020
Supplement Date Received by Manufacturer02/10/2021
Initial Report FDA Received Date08/07/2020
Supplement Report FDA Received Date02/10/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age25 YR
Patient Weight86
-
-