• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC SOFAMOR DANEK USA, INC CD HORIZON SPINAL SYSTEM; SCREWDRIVER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC SOFAMOR DANEK USA, INC CD HORIZON SPINAL SYSTEM; SCREWDRIVER Back to Search Results
Model Number 7570923
Medical Device Problem Code Material Fragmentation (1261)
Health Effect - Clinical Code Foreign Body In Patient (2687)
Date of Event 07/16/2020
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
Information was received from a healthcare professional (hcp) via a medtronic representative regarding a patient implanted with a screw using a driver for a spondylolisthesis l4-5.It was reported that driver tip broke off into the head of the screw.There was no patient symptoms or complications as a result of this event.The instrument broke and there was a fragment of the instrument remaining in the patient i.E.The tip of the broken driver is in the screw head.The pre- op diagnosis was spondylolisthesis l4-5.No additional treatment or surgery performed as a result of this event.The product was replaced by a medtronic product.No further complications were reported/ anticipated.
 
Additional Manufacturer Narrative
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Additional Manufacturer Narrative
D2: correction in product code medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Additional Manufacturer Narrative
B5: additional information received.H3: product analysis found that visually confirmed that the instrument tip of the quick connect end has been broken off, consistent with interface during usage.Inspection of the material hardness confirmed conformance to print specification.Fracture surface analysis of both instruments identified a fairly flat fracture surface and circular material movement, consistent with torsional overload.This is consistent with torsional overload.H6: analysis resulted in change in eval.Code method and eval.Code result.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
Information was received from a healthcare professional (hcp) via a medtronic representative regarding a patient implanted with a screw using drivers for a spondylolisthesis l4-5.It was reported tips of both drivers broke off into the head of the screw.There was no patient symptoms or complications as a result of this event.The instruments broke and there were fragments of the instruments remaining in the patient i.E.The tips of the broken drivers are in the screw head.The pre- op diagnosis was spondylolisthesis l4-5.No additional treatment or surgery performed as a result of this event.The products were replaced by medtronic products.No further complications were reported/ anticipated.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CD HORIZON SPINAL SYSTEM
Common Device Name
SCREWDRIVER
Manufacturer (Section D)
MEDTRONIC SOFAMOR DANEK USA, INC
4340 swinea rd
memphis TN 38118
MDR Report Key10377873
Report Number1030489-2020-01046
Device Sequence Number3450853
Product Code HXX
UDI-Device Identifier00613994236012
UDI-Public00613994236012
Combination Product (Y/N)N
Initial Reporter StateIL
Initial Reporter CountryUS
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Remedial Action Replace
Type of Report Initial,Followup,Followup,Followup
Report Date (Section B) 11/09/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number7570923
Device Catalogue Number7570923
Device Lot NumberRS18K038
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/15/2020
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/17/2020
Supplement Date Received by Manufacturer10/21/2020
10/21/2020
07/17/2020
Initial Report FDA Received Date08/07/2020
Supplement Report FDA Received Date10/26/2020
10/26/2020
11/10/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient Age72 YR
-
-