| Model Number |
MMT-1715KL |
| Medical Device Problem Code |
Mechanical Problem (1384)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
08/02/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).Currently it is unknown whether or not the device may have caused or contributed to the event as no product has been returned.The device will be returned for analysis and further information will follow once the analysis has been completed.No conclusion can be drawn at this time.
|
| |
|
Event or Problem Description
|
|
Information received by medtronic indicated that the insulin pump received critical pump error.Open book image was showing on screen.No harm requiring medical intervention was reported.The insulin pump will be returned for analysis.
|
| |
|
Additional Manufacturer Narrative
|
|
After battery installation, the device displayed a critical pump error (open book image) due to corrosion and mold around the battery connectors and the terminals of the keypad flex cable.Unable to perform the displacement, rewind, prime or seating, basic occlusion, force sensor, and occlusion tests due to the critical pump error.
|
| |
|