SMITH & NEPHEW, INC. LIGHT SOURCE 500XL XENON; IMAGE, ILLUMINATION, FIBEROPTIC, FOR ENDOSCOPE
|
Back to Search Results |
|
Model Number 72200568 |
Device Problem
Power Problem (3010)
|
Patient Problem
No Patient Involvement (2645)
|
Event Date 07/07/2020 |
Event Type
malfunction
|
Event Description
|
It was reported the light source presented a 13 lamp failure error and was making a clicking sound, then it turned off.No case reported; therefore, there was no patient involvement.All available information has been disclosed.If additional information should become available, a supplemental report will be submitted accordingly.
|
|
Manufacturer Narrative
|
The reported device, intended for use in treatment, was received for evaluation.There was a relationship found between the returned device and the reported incident.A visual inspection was performed on the exterior of product and no physical damage was observed.Complaint of e13 error was confirmed.Light source failed to ignite lamp and failed with an e13 temp error.Cause of ignition failure and error is a defective electronic component on the power supply.Unit powered up and passed functional testing with a known good power supply installed.The complaint investigation has concluded the cause of the failure to be a defective electronic component.A review of the device history records showed there were no indications to suggest that the product did not meet manufacturing specification or would not be able to perform as intended.A complaint history review concluded this was a repeat issue.
|
|