| Model Number |
82300 |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Codes |
No Known Impact Or Consequence To Patient (2692); No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
03/13/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Lot number and expiry are not available at this time.Investigation is in process.A follow up report will be provided.
|
| |
|
Event or Problem Description
|
|
The customer reported an occlusion and air bubbles during a platelet collection.It is unknown at this time the location of the air bubbles or if air was returned to the donor.Patient (donor) information and outcome are not available at this time.The platelet collection set is not available for return because it was discarded by the customer.
|
| |
|
Additional Manufacturer Narrative
|
|
Investigation: the dlogs were unavailable for this event.The customer provided information about a similar incident (1722028-2020-00363).The dlog for the similar incident indicated "possible air block" because at 21 minutes into the procedure, after receiving an alarm for low platelet concentration, the operator performed an air block recovery.It is possible, however not conclusive, the customer received a similar verification flag for this procedure.The customer did not respond to requests to provide the lot number for the disposable; therefore, a dhr search could not be conducted for this specific incident.All lots must meet acceptance criteria for release.Root cause: a definitive root cause could not be determined.Possible causes for air bubbles include but are not limited to: an occlusion in the disposable set due to manufacturing error, including solvent or molding occlusions; incomplete loading of the disposable set causing an occlusion in the tubing and/or an air block, which restricted the flow.Introduction of air into the anti-coagulant (ac), saline or inlet line during or after disposable prime (if prime solution bag runs dry and air from the bag is pulled into the tubing); a kink or an occlusion in the collect line.
|
| |
|
Event or Problem Description
|
|
Due to eu personal data protection laws, the patient information is not available from the customer.
|
| |
|