STRYKER GMBH ONE-THIRD TUBULAR PLATE VARIAX2 5 HOLE / L59MM; PLATE, FIXATION, BONE Back to Search Results
Model Number 626675
Medical Device Problem Code Material Twisted/Bent (2981)
Health Effect - Clinical Codes Failure of Implant (1924); Injury (2348)
Date of Event 07/13/2020
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Device will not be returned.If additional information becomes available, it will be provided in a supplemental report.Retained by hospital.
 
Event or Problem Description
It was reported that the patient's left radius was revised 1 month post-op due to a bent plate.The bent radius plate and an ulnar plate were revised.Rep confirmed there are no allegations against the ulnar plate.
 
Additional Manufacturer Narrative
The reported event could not be confirmed, since the device was not returned for evaluation and no other evidence was provided.The device inspection was not possible as the product was not returned for investigation.A review of the device history for the reported lot did not indicate any abnormalities.No corrective actions are required at this time.No indications of material, manufacturing or design related problems were found during the investigation.A review of the labeling did not indicate any abnormalities.The instructions for use states: ¿post-operative patient activity: these implants are neither intended to carry the full load of the patient acutely, nor intended to carry a significant portion of the load for extended periods of time.For this reason post-operative instructions and warnings to patients are extremely important.External immobilization (e.G.Bracing or casting) may be employed until x-rays or other procedures confirm adequate bone consolidation.The implant is intended for temporary bone fixation.In the event of a delay in bone consolidation, or if such consolidation does not take place, or if explantation is not carried out, complications may occur, for example fracture or loosening of the implant or instability of the implant system.Regular post-operative examinations (e.G.X-ray checks) are advisable.¿ more detailed information about the complaint event as well as the affected device must be available in order to determine the root cause of the complaint event.However, one of the probable causes of failure could be patient factor such as incorrect behavior and high level of activity after the implantation.If device is returned or any further information is provided, the investigation report will be reassessed.
 
Event or Problem Description
It was reported that the patient's left radius was revised 1 month post-op due to a bent plate.The bent radius plate and an ulnar plate were revised.Rep confirmed there are no allegations against the ulnar plate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ONE-THIRD TUBULAR PLATE VARIAX2 5 HOLE / L59MM
Common Device Name
PLATE, FIXATION, BONE
Manufacturer (Section D)
STRYKER GMBH
bohnackerweg 1
postfach
selzach 2545
SZ  2545
MDR Report Key10379524
Report Number0008031020-2020-02086
Device Sequence Number16936070
Product Code HRS
UDI-Device Identifier07613327088656
UDI-Public07613327088656
Combination Product (Y/N)N
Initial Reporter StateNC
Initial Reporter CountryUS
PMA/510(K) Number
K151879
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2020
Device Explanted Year2020
Reporter Type Manufacturer
Report Source health professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 10/09/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Expiration Date12/31/2023
Device Model Number626675
Device Catalogue Number626675
Device Lot NumberD13751
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 07/13/2020
Supplement Date Received by Manufacturer09/16/2020
Initial Report FDA Received Date08/07/2020
Supplement Report FDA Received Date10/09/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age15 YR
-
-