• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION DAKOTA; DISLODGER, STONE, BASKET, URETERAL, METAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION DAKOTA; DISLODGER, STONE, BASKET, URETERAL, METAL Back to Search Results
Model Number M0063905000
Medical Device Problem Code Tear, Rip or Hole in Device Packaging (2385)
Health Effect - Clinical Codes No Consequences Or Impact To Patient (2199); No Known Impact Or Consequence To Patient (2692)
Date of Event 07/12/2020
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).The device has not been received for analysis.Upon receipt and completion of the failure analysis of the complaint device, if there is any further relevant information from that review, a supplemental mdr will be filed.
 
Event or Problem Description
It was reported to boston scientific corporation that a dakota basket device was to be used in a retrograde intrarenal surgery procedure on (b)(6) 2020.According to the complainant, during preparation, a slit was noticed on the packaging.Another dakota basket was used to complete the procedure.There were no patient complications as a result of this event.
 
Additional Manufacturer Narrative
Block h6: device code 2385 captures the reportable event of slit in device packaging.Block h10: visual analysis found the device was returned in its original unopened pouch.The heat seal ran throughout the entire length of the pouch without any breakage.The pouch was found wrinkled in some areas.A cut was found on the back of the pouch that seemed to have been done with a sharp tool, which confirms the reported event.The customer had provided a photo of the device showing the damage as well.It is most likely that handling and manipulation of the device during transport and/or storage could have contributed to the encountered issue.Manipulation of the device can lead to damaging of the package during transport and/or storage.Therefore, the most probable root cause is cause traced to transport/storage.A review of the device history record (dhr) was performed and confirmed that this device met all material, assembly, and performance specifications at the time of release for distribution.
 
Event or Problem Description
It was reported to boston scientific corporation that a dakota basket device was to be used in a retrograde intrarenal surgery procedure on (b)(6) 2020.According to the complainant, during preparation, a slit was noticed on the packaging.Another dakota basket was used to complete the procedure.There were no patient complications as a result of this event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DAKOTA
Common Device Name
DISLODGER, STONE, BASKET, URETERAL, METAL
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
MDR Report Key10379644
Report Number3005099803-2020-03114
Device Sequence Number11236918
Product Code FFL
UDI-Device Identifier08714729937364
UDI-Public08714729937364
Combination Product (Y/N)N
Initial Reporter CountrySN
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Nurse
Type of Report Initial,Followup
Report Date (Section B) 09/24/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date10/03/2022
Device Model NumberM0063905000
Device Catalogue Number63057
Device Lot Number0024546054
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/12/2020
Initial Date Received by Manufacturer 07/12/2020
Supplement Date Received by Manufacturer08/31/2020
Initial Report FDA Received Date08/07/2020
Supplement Report FDA Received Date09/24/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-