|
Siemens has investigated the event of discordant low results with high sensitivity troponin i (tnih) on the dimension vista system with lot 20008bb.Siemens has observed a negative bias across the analytical measurement range of the tnih assay.Siemens tested lot 20008bb and a negative bias was confirmed.Siemens internal testing observed an average bias for patient samples for lot 20008bb of -20%.Urgent medical device correction, vc-20-03.A.Us dated july 24, 2020 and urgent field safety notice vc-20-03.A.Ous dated july 2020 is addressed to customers who had been shipped dimension vista® high sensitivity troponin i (tnih) lots 20008bb, 20035bc and 20135bb.Customer were instructed that they should recalibrate the tnih lots and to enter lot specific correlation factors c0 and c1 in the method configuration screen per flex reagent lot as listed in the communications.After applying correlation factors to the specified lot, product claims listed in the instructions for use are met.Siemens is working to restore assay performance, and until such time, subsequent lots may contain an alert card in the carton containing lot specific correlation factors.
|