• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: PHILIPS ULTRASOUND, INC S8-3T MICRO TEE TRANSDUCER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

PHILIPS ULTRASOUND, INC S8-3T MICRO TEE TRANSDUCER Back to Search Results
Model Number 989605420183
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Return of the suspect transducer is anticipated.Evaluation of the transducer will be included in a follow up report upon its return and investigation completion.
 
Event or Problem Description
A customer reported an s8-3t model transducer had an image quality issue.There was no injury associated with this event.
 
Additional Manufacturer Narrative
Evaluation of the returned s8-3t model transducer confirmed the poor image quality issue as described by the customer.Investigation of the device revealed oil separation in the window, chemical damage to the sheath, dented distal and proximal rubber, a scratched handle, the articulation would not properly move backwards, a scratched strain relief at the connector¿s end, the connector housing had chipped paint, and the connector had debris.The extensive physical damage to the transducer inhibited the overall performance of the device and is indicative of improper maintenance.Submission of a report does not constitute an admission that medical personnel, user facility, importer, distributor, manufacturer, or product caused or contributed to the event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
S8-3T MICRO TEE TRANSDUCER
Common Device Name
S8-3T MICRO TEE TRANSDUCER
Manufacturer (Section D)
PHILIPS ULTRASOUND, INC
22100 bothell everett highway
bothell WA 98021
MDR Report Key10379969
Report Number3019216-2020-00051
Device Sequence Number16894449
Product Code ITX
UDI-Device Identifier00884838067523
UDI-Public(01)00884838067523
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
K030455
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source health professional
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 07/22/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number989605420183
Device Lot NumberB37JH7
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/27/2020
Was the Report Sent to FDA? No
Initial Date Received by Manufacturer 07/22/2020
Supplement Date Received by Manufacturer07/22/2020
Initial Report FDA Received Date08/07/2020
Supplement Report FDA Received Date02/23/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-