| Model Number |
MMT-1780KL |
| Medical Device Problem Code |
Device Difficult to Program or Calibrate (1496)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
08/03/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).Currently it is unknown whether or not the device may have caused or contributed to the event as no product has been returned.The device will be returned for analysis and further information will follow once the analysis has been completed.No conclusion can be drawn at this time.
|
| |
|
Event or Problem Description
|
|
Information received by medtronic indicated that the insulin pump had stuck button alarm.Customer stated that they did not press a button down for 3 minutes or longer.No harm requiring medical intervention was reported.The device will be returned for analysis.
|
| |
|
Additional Manufacturer Narrative
|
|
Device received with fully functional keypad.Peeled off keypad overlay and no damage noted on keypad traces.Keypad flex connector is plugged in properly.Device properly tested with displacement test and self test.No button error alarm noted upon testing.(b)(4).
|
| |
|