PENUMBRA, INC. BENCHMARK 6F 071 DELIVERY CATHETER; DQY Back to Search Results
Model Number BMK6F95BER120
Medical Device Problem Codes Break (1069); Unraveled Material (1664); Physical Resistance/Sticking (4012)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 06/28/2020
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
This device is available for return.A follow up mdr will be submitted upon completion of the device investigation.
 
Event or Problem Description
The patient was undergoing a coil embolization procedure in the posterior communication artery (pcom) using a benchmark 6f 071 delivery catheter (benchmark) and a non-penumbra 6f radial sheath.During the procedure, the physician inserted the benchmark through the sheath and into the patient using the right radial access.While advancing up the benchmark towards the target vessel, the physician experienced resistance and decided to remove the benchmark.Upon removal of the benchmark from the sheath, the physician noticed that the tip of the benchmark had broken off.It was reported that the braiding of the benchmark unraveled within the axillary and right subclavian arteries, and that 10 centimeters of the tip of the benchmark was in the radial artery.Subsequently, an angiogram revealed patent subclavian, axillary, brachial and ulnar arteries, but a fully occluded radial artery.The physician then made a second puncture in the groin and completed the coiling.Afterwards, the broken benchmark tip was removed using a snare device.The final fragment of the benchmark was removed via surgery in the right axillary region.The procedure ended at this point.There was no report of an adverse effect to the patient.
 
Additional Manufacturer Narrative
Results: the returned benchmark was stretched approximately 46.0 cm, 47.0 cm, 48.0 cm and 52.0 cm from the hub.The device was fractured approximately 53.0 cm, 59.0 cm, 77.0 cm and 92.0 cm from the hub.The total length of the returned benchmark was 92.0 cm.Conclusions: evaluation of the returned benchmark confirmed a fracture and revealed stretching near the site of the fracture.If the device is forcefully retracted against resistance, the catheter may begin to stretch.If the device continues to be retracted against resistance it may subsequently fracture as a result.Further evaluation of the device revealed additional fractures on the distal portion of the fractured catheter.This damage was likely incidental to the complaint and may have occurred during attempts to remove the device from the patient.Penumbra catheters are inspected during in-process inspection and during quality inspection after manufacturing.The manufacturing records for this lot were reviewed and did not reveal any outstanding discrepancies, design, or quality concerns.H3 other text : placeholder.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BENCHMARK 6F 071 DELIVERY CATHETER
Common Device Name
DQY
Manufacturer (Section D)
PENUMBRA, INC.
one penumbra place
alameda CA 94502
MDR Report Key10380178
Report Number3005168196-2020-01138
Device Sequence Number3669743
Product Code DQY
UDI-Device Identifier00814548014128
UDI-Public00814548014128
Combination Product (Y/N)Y
Initial Reporter CountryUK
PMA/510(K) Number
K142321
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source foreign,user facility
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 07/15/2020
Report Date (Section F) 01/01/2005
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date06/02/2022
Device Model NumberBMK6F95BER120
Device Catalogue NumberBMK6F95BER120
Device Lot NumberF90707
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/09/2020
Was the Report Sent to FDA? Yes
Date Report to Manufacturer01/10/2005
Initial Date Received by Manufacturer 07/15/2020
Supplement Date Received by Manufacturer09/28/2020
Initial Report FDA Received Date08/07/2020
Supplement Report FDA Received Date09/30/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age50 YR
Date Report Sent to FDA01/01/2005
-
-