MDT POWERED SURGICAL SOLUTIONS LEGEND DISSECTING TOOL; DRILLS, BURRS, TREPHINES & ACCESSORIES (SIMPLE, POWERED) Back to Search Results
Model Number 8TA11
Medical Device Problem Code Fracture (1260)
Health Effect - Clinical Codes Foreign Body In Patient (2687); Device Embedded In Tissue or Plaque (3165)
Date of Event 07/10/2020
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
No conclusion can be drawn.No evaluation was performed, as the device was not returned.If the device is returned in the future, product analysis may be performed.The user manual contains the following warning ¿do not use excessive pressure, such as bending or prying, on attachments or dissecting tools.This may cause tool to bend or break and cause injury to patient, operator and/or operating room staff.¿ we will continue to track and trend this complaint type.If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
It was reported that 2 tools were broken during the procedure.It was also added that the broken pieces were retrieved from the patient and the procedure was completed using a back-up device with no delay.
 
Additional Manufacturer Narrative
Report confirmed.Evaluation determined that the tools were broken.The likely cause was identified as the distal bearing from the attachment that was used with the tool is no longer supporting the tool resulting in the tool contacting the inside of the attachment housing.This tool contact may cause the attachment to overheat and fracture the tool at the shoulder when presented with sufficient side loading.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
On follow up, it was confirmed that there was no patient impact and no further information can be provided regarding the event.
 
Additional Manufacturer Narrative
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
On follow up, it was reported that the em200 will not be sent for analysis.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LEGEND DISSECTING TOOL
Common Device Name
DRILLS, BURRS, TREPHINES & ACCESSORIES (SIMPLE, POWERED)
Manufacturer (Section D)
MDT POWERED SURGICAL SOLUTIONS
4620 n beach st
fort worth TX 76137
MDR Report Key10380210
Report Number1625507-2020-00118
Device Sequence Number1730183
Product Code HBE
UDI-Device Identifier00643169765689
UDI-Public00643169765689
Combination Product (Y/N)N
Initial Reporter CountryBE
PMA/510(K) Number
K020069
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source foreign,health professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date (Section B) 12/22/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date04/29/2024
Device Model Number8TA11
Device Catalogue Number8TA11
Device Lot Number0217584460
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/28/2020
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/16/2020
Supplement Date Received by Manufacturer09/15/2020
12/21/2020
Initial Report FDA Received Date08/07/2020
Supplement Report FDA Received Date10/15/2020
12/22/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
EM200 (S/N: (B)(6)); MR7 (UNKNOWN SERIAL NUMBER); MR7 (UNKNOWN SERIAL NUMBER)
Outcome Attributed to Adverse Event Other;
-
-