• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZOLL MANUFACTURING CORPORATION LIFEVEST WCD 4000 SYSTEM; WEARABLE CARDIOVERTER DEFIBRILLATOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZOLL MANUFACTURING CORPORATION LIFEVEST WCD 4000 SYSTEM; WEARABLE CARDIOVERTER DEFIBRILLATOR Back to Search Results
Model Number WCD 4000
Device Problem Connection Problem (2900)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 06/06/2020
Event Type  malfunction  
Manufacturer Narrative
Device evaluation of battery charger/modem sn (b)(4) has been completed.The reported problem (charger power connector problem) has been confirmed.Upon investigation the battery charger/modem was unable to power on.The cause for the failure was isolated to a damaged power supply connector.The root cause for the damaged power supply connector was excessive force.No adverse event resulted from the defective battery charger/modem.
 
Event Description
A us distributor reported that a patient had a battery charger power connector problem.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LIFEVEST WCD 4000 SYSTEM
Type of Device
WEARABLE CARDIOVERTER DEFIBRILLATOR
Manufacturer (Section D)
ZOLL MANUFACTURING CORPORATION
121 gamma drive
pittsburgh, pa
Manufacturer (Section G)
ZOLL MANUFACTURING CORPORATION
121 gamma drive
pittsburgh, pa
Manufacturer Contact
matthew saich
121 gamma drive
pittsburgh, pa 
9683333
MDR Report Key10380432
MDR Text Key203010100
Report Number3008642652-2020-06586
Device Sequence Number1
Product Code MVK
UDI-Device Identifier00855778005029
UDI-Public00855778005029
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P010030
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type distributor
Reporter Occupation Other
Type of Report Initial
Report Date 07/29/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/07/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Model NumberWCD 4000
Device Catalogue Number10A0985
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/02/2020
Date Manufacturer Received07/13/2020
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured07/31/2018
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-