• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION NC EMERGE; CATHETERS, TRANSLUMINAL CORONARY ANGIOPLASTY, PERCUTANEOUS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION NC EMERGE; CATHETERS, TRANSLUMINAL CORONARY ANGIOPLASTY, PERCUTANEOUS Back to Search Results
Model Number 7214
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Patient Involvement (2645)
Date of Event 07/14/2020
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that shaft break occurred.A 2.50mm x 20mm nc emerge balloon catheter was selected for use.However, when loading onto the wire, the body of the balloon broke.It was noted that the wire was bent prior to insertion and it snapped when attempted to strenghten it out.
 
Additional Manufacturer Narrative
Device evaluated by mfr: returned product consisted of an nc emerge balloon catheter.The shaft, hypotube, tip and balloon were microscopically and visually examined.There were numerous kinks.There was a complete separation at 54.3cm distal of the strain relief.The balloon was tightly folded.Inspection of the remainder of the device presented no other damage or irregularities.Product analysis confirmed the reported broken shaft as the hypotube was separated.
 
Event or Problem Description
It was reported that shaft break occurred.A 2.50mm x 20mm nc emerge balloon catheter was selected for use.However, when loading onto the wire, the body of the balloon broke.It was noted that the wire was bent prior to insertion and it snapped when attempted to strenghten it out.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NC EMERGE
Common Device Name
CATHETERS, TRANSLUMINAL CORONARY ANGIOPLASTY, PERCUTANEOUS
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
two scimed place
maple grove MN 55311
MDR Report Key10380595
Report Number2134265-2020-09846
Device Sequence Number16618526
Product Code LOX
UDI-Device Identifier08714729846635
UDI-Public08714729846635
Combination Product (Y/N)N
Initial Reporter StateMO
Initial Reporter CountryUS
PMA/510(K) Number
K141236
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 09/10/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date03/09/2022
Device Model Number7214
Device Catalogue Number7214
Device Lot Number0025325206
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/28/2020
Initial Date Received by Manufacturer 07/16/2020
Supplement Date Received by Manufacturer08/19/2020
Initial Report FDA Received Date08/07/2020
Supplement Report FDA Received Date09/10/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age82 YR
-
-