| Model Number |
7214 |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
No Patient Involvement (2645)
|
| Date of Event |
07/14/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that shaft break occurred.A 2.50mm x 20mm nc emerge balloon catheter was selected for use.However, when loading onto the wire, the body of the balloon broke.It was noted that the wire was bent prior to insertion and it snapped when attempted to strenghten it out.
|
| |
|
Additional Manufacturer Narrative
|
|
Device evaluated by mfr: returned product consisted of an nc emerge balloon catheter.The shaft, hypotube, tip and balloon were microscopically and visually examined.There were numerous kinks.There was a complete separation at 54.3cm distal of the strain relief.The balloon was tightly folded.Inspection of the remainder of the device presented no other damage or irregularities.Product analysis confirmed the reported broken shaft as the hypotube was separated.
|
| |
|
Event or Problem Description
|
|
It was reported that shaft break occurred.A 2.50mm x 20mm nc emerge balloon catheter was selected for use.However, when loading onto the wire, the body of the balloon broke.It was noted that the wire was bent prior to insertion and it snapped when attempted to strenghten it out.
|
| |
|