Model Number 71175077
Medical Device Problem Code Degraded (1153)
Health Effect - Clinical Code No Patient Involvement (2645)
Date of Event 06/12/2020
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that before surgery while getting ready for the case the screw driver was found to be stripped.It is unknown if there was a delay o how the procedure was finished.No injuries reported at this time.
 
Additional Manufacturer Narrative
The device, intended use in treatment, was returned for evaluation.A visual inspection confirmed the tip is damaged on the evos small t8 fixed handle linear driver, which would cause the device not to function as intended.The device shows significant signs of wear/usage.A review of the manufacturing records did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.At this time, we have no reason to suspect that the products failed to meet any product specifications at the time of manufacture.A complaint history review found related failures; this failure mode will be monitored for future complaints for any necessary corrective actions.Damage from prolonged use, misuse or rough handling are likely probable causes of the reported event.We recommend that all reusable instruments be routinely inspected for wear and damage and replaced as necessary.This is a reusable instrument that can be exposed to numerous surgeries; damage from repeated use can occur.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EVOS SMALL T8 FIXED HANDLE LINEAR DRIVER
Common Device Name
SCREWDRIVER
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
MDR Report Key10380674
Report Number1020279-2020-03783
Device Sequence Number3551316
Product Code HXX
UDI-Device Identifier00885556636077
UDI-Public00885556636077
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 08/25/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number71175077
Device Catalogue Number71175077
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/03/2020
Initial Date Received by Manufacturer 06/12/2020
Supplement Date Received by Manufacturer08/14/2020
Initial Report FDA Received Date08/07/2020
Supplement Report FDA Received Date08/25/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-