JOHNSON & JOHNSON SURGICAL VISION, INC. IDESIGN AWS SYSTEM; REFRACTIVE MEASUREMENT Back to Search Results
Model Number 0110-3450
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Therapeutic Effects, Unexpected (2099)
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Date of event: unknown, not provided.A record review was performed.A product deficiency review was performed and there is no product deficiency identified.A document, service history, and trending was reviewed.There is not a recognizable adverse trend.The risks and mitigations associated with the complaint issue are identified in existing risk documents and no new risks were identified as part of this investigation.A labeling review was conducted; the operator manual for the system was reviewed and found to include adequate instructions for use, warnings and operational errors.The review of the device history record (dhr) for idesign aws system showed that there were no issues or non-conformities.The system and its components met all specifications prior to being released.Manufacturing has been ruled out as a potential cause for the reported issue.Based on the investigation results, no corrective action has been issued.Based on the investigation results there is no indication of a product quality deficiency.Johnson & johnson surgical vision will continue to monitor this type of complaints.All pertinent information available to johnson & johnson surgical vision, inc has been submitted.All pertinent information available to johnson & johnson surgical vision, inc has been submitted.
 
Event or Problem Description
The surgery center reported a case of over-correction following idesign ilasik treatment.A re-treatment was performed and the patient is doing well post-op.There was no loss of best corrected visual acuity (bcva) reported.This report is for idesign aws system.A separate report will be submitted for the excimer laser system.
 
Additional Manufacturer Narrative
This correction mdr is being submitted to address inaccurate information in section f on the xml version of the previous submission caused by a software glitch in the transmission process.The internal non-conformance numbers for this software issue are nr-0148919 and capa-010215.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
IDESIGN AWS SYSTEM
Common Device Name
REFRACTIVE MEASUREMENT
Manufacturer (Section D)
JOHNSON & JOHNSON SURGICAL VISION, INC.
1700 e st andrew place
santa ana CA 92705
MDR Report Key10382161
Report Number3006695864-2020-00372
Device Sequence Number1896573
Product Code HKO
UDI-Device Identifier05050474651432
UDI-Public(01)05050474651432
Combination Product (Y/N)Y
Initial Reporter StateMO
Initial Reporter CountryUS
PMA/510(K) Number
K000327
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source health professional,user faci
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 10/29/2020
Report Date (Section F) 01/01/2005
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model Number0110-3450
Device Catalogue Number0110-3450
Was Device Available for Evaluation? Yes
Was the Report Sent to FDA? No
Date Report to Manufacturer01/10/2005
Initial Date Received by Manufacturer 07/14/2020
Supplement Date Received by Manufacturer10/25/2020
Initial Report FDA Received Date08/07/2020
Supplement Report FDA Received Date10/29/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
EXCIMER, SERIAL NUMBER (B)(6)
Outcome Attributed to Adverse Event Required Intervention;
Date Report Sent to FDA01/01/2005
-
-